Support for the 85033 SMA calibration and verification kit.

Outside Kirkby Microwave SMA calibration kit Inside Kirkby Microwave SMA calibration kit

Contents

This page describes the 85033 kit in detail. Other pages on our site show how to set up calibration kits for various VNAs. Unless you know how to set up a calibration kit, you will need to read both sections to familirize yourself with this kit, and also how to set it up.

  1. Introduction.
  2. Frequency range.
  3. Supported instruments. Virtually all VNAs can be made to work with the 85033.
  4. Identifying standard components of the SMA kit.
  5. Identifying optional components of the SMA kit.
  6. Calibration kit coefficients.
  7. Configuring the VNA to use the calibration kit.
  8. Using the optional (BUT HIGHLY RECCOMENDED) torque wrench. This is option 004
  9. Calibration.
  10. Verification. - How to verify the VNA is correctly configured to use the N calibration kit.
  11. Using the optional reverse SMA adapters (option 001)
  12. Using the optional U.FL adapters and cable (option 002)

1. Introduction.

The Kirkby Microwave 85033 SMA calibration and verification kit is our most popular kit. It is designed to be used with virtually any vector network analyzer (VNA) to allow the VNA to be calibrated by the user. It is designed to fill a gap in the market between high end calibration kits from companies like Keysight, Anritsu, Rosenberger, Maury Microwave etc, and some very low cost kits often sold on eBay by people who assemble an open, short and 50 Ohm load, and call it a calibration kit. There's a review by Dr. R. Watson at Bath University of this kit. We invite you to read that.

The kit is particularly popular with

2. Frequency Range

The standard frequency range of the 85033 SMA calibration and verification kit is DC-7 GHz. The loads will have a return loss of at least 32 dB to 7 GHz, and the phase of the calibration standards will follow the mathematical model within the accuracy given in the specifications. We can produce kits to 8, 9, 10 11 or 12 GHz, at increasing cost. These have higher frequency loads and other changes which make operation at higher frequencies possible. We can also give calibration data at user requested frequencies - e.g. 10.368 GHz is commonly requested. Please contact us about this for the price and availability.

3. Supported Instruments.

VNAs manufactured by the following companies are fully supported.

We may be able to support other makes, such as Advantest, but we would need to be made aware of the model number first, to allow us to research this.

It is necessary to load the coefficients of the kit into the VNA. In some instances this will need to be done from the front panel, but for the following VNAs we can supply files which make uploading very easy

You will not find files for all these instruments on our website, but we can create them for you, with the exception of some old Rohde & Schwarz VNAs and also some old Anritsu VNAs. We can create calibration kit files for all current instruments.

4. Identifying standard components, supplied with all SMA calibration kits.

The 6 most important parts are the two opens, two shorts, and two load standards. (Rohde & Schwarz refer to the load as the 'match' standard, but we will use the term 'load' as it is more commonly used.)

open and short VNA calibration standards SMA male load SMA male load

To allow devices to be connected which don't have the appropiate sex of connectors for the device under test, then we include two adapters. One is a male-male thru and the other a female-female thru. (Although it is not obvious from the photograph, the female-female thru has flats on it, which allow it to be held with a 5.5 mm spanner.

female-female SMA calibrated adapter female-female SMA calibrated adapter

If your VNA supports unknown thru (Agilent Keysight term) or UOSM (Rohde and Schwarz term), then you will not need to worry about the delay of these adapters, but if the VNA does not support these method of calibration, then you need to consider the delay (ps) or electrical length (mm) of the adapters, which will be supplied.

For verification purposes, there's also an attenuator with measured S-parameters, which can help verify the system is set up correctly later. Attenuator with S-parameters measured.

5. Identifying the optional components supplied at extra cost.

The kits can be provided with optional extras, such as reverse SMA adapters (option 001), U.FL adapters (option 002) or a torque wrench (option 004). A picture of the torque wrench is shown

Huber and Suhner torque wrench

This provides a fixed torque of approximately 4 lb inch, which is ideal for all SMA connectors.

The kit can also be supplied with a set of two reverse SMA adapters, which allow one to connect a standard SMA connector to either a reverse male or reverse female SMA connector.

Male SMA to female RP. SMA adapter Female SMA to male RP. SMA adapter

6. Coefficients of the calibration kit

Since late 2017, each kit has been supplied with individual data for that kit, which is put on the USB stick. Kits sold prior to this have generic coefficients. If you are unsure of the coefficients of your kit, please provide us the serial number of the kit. Coefficients are supplied in two forms

1. Human readable text files.

Using one of these files, it is possible to configure almost any VNA to work with our calibration kits. Anritsu-coefficients.txt
Copper-Mountain-coefficients.txt
HP-Agilent-Keysight-coefficients.txt
L_A_Techniques-coefficients.txt
PicoVNA-coefficients.txt
Rohde-and-Schwarz-coefficients.txt
Tektronix-coefficients.txt

2. Files that can be read by specific models of VNAs.

We provide files readable by some Anritsu, Agilent and Keysight VNAs with every kit, as we can generate these automatically, without any human intervention, in about 1 second. These include

For other instruments, like those from Copper Mountain, Rohde & Schwarz, HP/Agilent, Pico Technology etc, we will generate files readable by a VNA when a customer requests them.

Despite more of our customers own HP 8753s than any other VNA, we have no completely automated way of generating files for the 8753s. We need to use some software Agilent wrote, or an actual 8753ES, since the file is a binary file in a proprietary binary format.

7. Configuring the VNA to use the calibration kit.

The following shows some pages we have on configuring specific VNAs. With the majority of customers owning instruments from Anritsu, HP/Agilent/Keysight, Rohde & Schwarz as well as Copper Mountain, we have more experience with them. But we do have customers with some of the rarer VNAs.

Anritsu
Copper Mountain (1-port and transmission/reflection)
Copper Mountain (Full 2-port & 4-port)
HP 8714
HP/Agilent 8753
HP/Agilent 8719, 8720 & 8722
HP/Agilent 8510
Agilent/Keysight PNA
Agilent/Keysight PNA-X
Agilent/Keysight FieldFox
Agilent/Keysight ENA
LA Techniques
Rohde & Schwarz

8. Using the optional torque wrench - this is option 004.

Using the correct torque on SMA connectors is particularly important, as hand tight is too lose, and yet it is easy to over-tighten with a spanner, which will either cause immediate failure, or reduce the lifetime. (For N-connectors, the use of torque wrench is not critical, as hand tight is really good enough.) The use of a torque wrench on SMA improves measurement repeatability, especially at the higher frequencies.

The torque wrench should only be used to tighten the SMA connectors - never to losen them. For losening the connector, an 8 mm spanner should be used.

Using an SMA torque wrench

  1. Ensure the correct torque wrench is used. The ones supplied by Kirkby Microwave are 4 lb inch, which is suitable for any SMA connector. Althought SMA and 3.5 mm are mechanically compatible, a torque wrench designed for 3.5 mm connectors will have a torque of 8 lb inch, which is far too high for SMA connectors.
  2. Position the torque wrench and a second open-ended spanner to hold the device or cable. For female thrus, a 5.5 mm thin spanner is needed, but for all other connections, an 8 mm spanner is needed. Apply them within 90° of each other. As the connection is tightened, they will get closer. If the nut has to be rotated such that the sepparation will exceed 90 degrees, then reposition the torque wrench. Ensure you support the devices to avoid putting stress on the SMA connectors.
  3. Hold the torque wrench lightly at the end of handle - do not grash the whole handle with the fist. Apply a force perpendicular to the torque wrench handle. Tighten only until the break point of the torque wrench is reached.

9. How to calibrate the VNA using the SMA calibration and verification kit.

10. How to verify the VNA is correctly configured by using the verification attenuator.

Once the VNA is set up to use the 85033 SMA calibration and verification kit, it is tempting to start making measurements. However, it is wise to verify that that the kit is correctly configured, and the VNA giving sensible results. Our kits include an attenuator

attenuator

which has been measured using an Agilent 85052B calibration kit. The results of the measurements are stored on a USB stick included with the kit. We have a page about verification in our FAQ.